Cranial Nerve Stimulator


e-TNS (External Trigeminal Nerve Stimulation)
s-TNS (Supraorbital Transcutaneous Nerve Stimulation)

Migraine treatment

Cefaly
Cefaly 2 European Model Cefaly 2 US Model
Cefaly Cefaly US
Cefaly 1 European Model Cefaly 1 US Model
Cefaly Cefaly US
  • Current Generator for maximum skin impedance of 2.2 KΩ
  • Automatic treatment interruption and signal when circuit impedance higher 2.2 KΩ
  • 3 standard programs :
    1. 100 Hz ; ramp up 14’ ; Imax 16 mA
    2. 60Hz ; ramp up 14’ ; Imax 16 mA
    3. 120Hz ; ramp up 12’ ; Imax 14 mA
  • Fully programmable : d = 10 to 500 μS ; F = 1 to 150 Hz ; I = 1 to 20 mA
  • Ramp up and session duration fully adjustable
  • 1 standard program :
    • 60Hz ; ramp up 14’ ; Imax 16 mA
www.cefaly.com www.cefaly.us

e-ONS (External Occipital Nerve Stimulation or Suboccipital Transcutaneous Nerve Stimulation

Occipital neuralgia

Cefaly
Cefaly Cefaly
  • Current Generator for maximum skin impedance of 2.2 KΩ
  • Automatic treatment interruption and signal when circuit impedance higher 2.2 KΩ
  • 3 standard programs :
    1. 100 Hz ; ramp up 14’ ; Imax 16 mA
    2. 60Hz ; ramp up 14’ ; Imax 16 mA
    3. 120Hz ; ramp up 12’ ; Imax 14 mA
  • Fully programmable : d = 10 to 500 μS ; F = 1 to 150 Hz ; I = 1 to 20 mA ;
  • Ramp up and session duration fully adjustable

Interface, software, accessories

PC connection cable

Cefaly connected
cefaly compliance
cefaly programmable

Transcranial Neurostimulator

t-DCS

The CEFALY t-DCS is small, portable and ergonomic. It is conceived to be easily used by patients at home. Currently only available for clinical investigation.

t-DCS
  • Current Generator for maximum skin impedance of 2.2 KΩ
  • Automatic treatment interruption and signal when circuit impedance higher 2.2 KΩ
  • 1 standard program :
    • I = 2 mA ; ramp up 30’’, session duration 20’ ; ramp down 30’’
  • Fully programmable with specific parameters on demand

Biofeedback

The CEFALY BIOFEED records the action potential of the head muscles through an electrode placed on the forehead. There is a feedback mechanism which is both visual and auditory. The visual signal comes from the Light Emitting Diode (LED) on the device. The auditory signal is composed of very audible beeps: the frequency of the beeps decrease with the muscles’ action potential. So the patient can hear that he/she is relaxing as the beeps increase in interval.

acquisition et analyse de biosignaux
Biofeed